|
|
|
LH-Irma
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Reagents | 100 tests kit | colour code | reconstitution |
| Coated tubes with anti-LH (monoclonal antibodies) | 2x50
|
ready to use
|
|
| Tracer: 125Iodine labelled anti-LH in TRIS-HCl buffer with BSA, azide (<0.1%) and inert red dye | 1vial 5.0 ml 700kBq | red | ready to use |
| standard zero in bovine serum with thymol | 1 vial lyoph. | yellow | add 2 ml distilled water |
| standard 1 to 6 (see exact values on labels) in bovine serum with thymol | 6 vials lyoph. | yellow | add 1 ml distilled water |
| wash solution conc. (TRIS-HCl) | 1 vial 10 ml | brown | dilute 70x with distilled water (use a magnetic stirrer) |
| control 1&2 in human serum with thymol | 2 vials lyoph. | silver | add 0.5 ml distilled water |
Note:
1. Use the zero
calibrator for sera dilutions.
2.
1 mIU of the calibrator preparation is equivalent to 1 mIU of 2nd
IRP 80/552.
5.
Supplies not provided
The
following material is required but not provided in the kit:
1.
Distilled water
2.
Pipettes for delivery of: 50 μl, 100 µl, 500 μl, 1 ml and 2
ml. (the use of accurate pipettes with disposable plastic tips is recommended)
3.
Vortex mixer
4.
Magnetic stirrer
5.
Tube shaker
6.
5 ml automatic syringe (Cornwall type) for washing
7.
Aspiration system (optional).
8.
Any gamma counter capable of measuring 125I may be used
(minimal yield 70%).
6.
Reagent preparation
A.
Calibrators
: Reconstitute the zero calibrator with 2 ml distilled water and the other
calibrators with 1 ml distilled water..
B.
Controls
: Reconstitute the controls with 0.5 ml distilled water.
C.
Working Wash solution
: Prepare an adequate volume of Working Wash solution by adding 69 volumes of
distilled water to 1 volume of Wash Solution (70x). Use a magnetic stirrer to
homogenize. Discard unused Working Wash solution at the end of the day.
7.
Storage ans expiration dating of reagents
-
Before opening or reconstitution, all kit components are stable until the
expiry date, indicated on the label, if kept at 2 to 8°C.
-
After reconstitution, calibrators and controls are stable for 3 days at
2-8°C.
For longer storage periods, aliquots should be made and kept at –20°C
for maximum 3 months. Avoid
subsequent freeze-thaw cycles.
-
Freshly prepared Working Wash solution should be used on the same day.
- After its first use, tracer
is stable until expiry date, if kept in the original well-closed vial at 2 to 8°C.
- Alterations in physical
appearance of kit reagents may indicate instability or deterioration.
8.
Specimen collection and preparation
Serum must be kept at 2 –
8°C.
If the test is not run within 24 hours, storage at –20°C
is recommended.
Avoid subsequent freeze-thaw cycles.
9. Procedure
A.
Handling notes
Do
not use the kit or components beyond expiry date. Do not mix materials from different kit lots.
Bring all the reagents to room temperature prior to use.
Thoroughly
mix all reagents and samples by gentle agitation or swirling. In order to avoid
cross-contamination, use a clean disposable pipette tip for the addition of each
reagent and sample.
High
precision pipettes or automated pipetting equipment will improve the precision.
Respect the incubation times.
Prepare a calibration
curve for each run, do not use data from previous runs.
B.
Procedure
1.
Label
coated tubes in duplicate for each calibrator, control and sample.
For
determination of total counts, label 2 normal tubes.
2.
Briefly
vortex
calibrators, controls and samples and dispense 100 μl
of each into the respective tubes.
3.
Dispense
50 μl
of anti-LH-125I tracer into each tube, including the uncoated tubes
for total counts.
4.
Shake
the rack containing the tubes gently by hand to liberate any trapped air
bubbles.
5.
Incubate
for 1 hour at room temperature on a tube shaker (700 rpm).
6.
Aspirate
(or decant) the content of each tube (except total counts).
Be sure that the plastic tip of the aspirator reaches the bottom of the
coated tube in order to remove all the liquid.
7.
Wash
the tubes with 2 ml Wash Solution (except total counts). Avoid foaming during the addition of the Working Wash
Solution.
8.
Aspirate
(or decant) the content of each tube (except total counts).
9.
Wash
again the tubes with 2 ml Wash Solution (except total counts) and aspirate (or
decant).
10.
After
the last washing, let the tubes standing upright for two minutes and aspirate
the remaining drop of liquid.
11.
Count
the tubes in a gamma counter for 60 seconds.
10. Calculation of results
1.
Calculate the mean of duplicate determinations.
2.
On semi logarithmic or linear graph paper plot the c.p.m. (ordinate) for
each calibrator against the corresponding concentration of LH (abscissa) and
draw a calibration curve through the calibrator points, reject the obvious
outliers.
3.
Read the concentration for each control and sample by interpolation on
the calibration curve.
4.
Computer assisted data reduction will simplify these calculations. If automatic result processing is used, a 4-parameter
logistic function curve fitting is recommended.
11.
Typical data
The
following data are for illustration only and should never be used instead of the
real time calibration curve.
|
LHsp-IRMA |
cpm |
B/T
(%) |
|
|
Total
count |
246440 |
100 |
|
|
Calibrator |
0.0
mIU/ml 1.8
mIU/ml 3.5
mIU/ml 9.9
mIU/ml 30.0
mIU/ml 97.0
mIU/ml 194.0
mIU/ml |
207 533 1165 3821 12920 50053 93732 |
0.1 0.2 0.5 1.6 5.2 20.3 38.0 |
12.
Performance and limitations
A.
Detection Limit
Twenty
zero calibrators were assayed along with a set of other calibrators. The
detection limit, defined as the apparent concentration two standard deviations
above the average counts at zero binding, was 0.2 mIU/ml.
B.
Specificity
Cross-reactive
hormones were added to a low and to a high LH value calibrator.
The apparent
LH response was measured.
|
added
Hormone |
LHsp
CAL 1 mIU/ml |
LHsp
CAL 5 mIU/ml |
|
- |
1.8 |
97 |
|
FSH
300 mIU/ml |
1.6 |
87 |
|
hCG
300000 mIU/ml |
2.4 |
100 |
|
TSH
300 µIU/ml |
3.6 |
85 |
C.
Precision
|
INTRA
ASSAY |
INTER
ASSAY |
||||||
|
Serum |
N |
<X>
±
S.D. mIU/ml) |
CV (%) |
Serum |
N |
<X>
±
S.D. (mIU/ml) |
CV (%) |
|
A B |
10 10 |
6.6
±
0.3 49.6
±
0.7 |
3.9 1.4 |
C D |
20 20 |
5.9
±
0.5 56.7
±
1.9 |
8.0 3.4 |
D.
Accuracy
RECOVERY TEST
|
Sample |
Added
LH (mIU/ml) |
Recovered
LH (mIU/ml) |
Recovery (%) |
|
1 2 |
0.5 1.5 5 14 46 0.5 1.5 5 14 46 |
0.7 1.8 4.5 12.7 46.1 0.7 1.3 4.7 12.9 43.8 |
130 117 89 90 100 140 87 94 92 95 |
|
DILUTION TEST |
|||
|
Sample |
Dilution |
Theoretical
Concent. (mIU/ml) |
Measured
Concent. (mIU/ml) |
|
1 2 |
1/1 1/2 1/4 1/8 1/16 1/32 1/1 1/2 1/4 1/8 1/16 1/32 |
- 39.9 19.9 10.0 5.0 2.5 - 60.5 30.3 15.1 7.6 3.8 |
79.7 41.1 19.9 10.1 5.3 2.2 121.0 53.0 27.9 14.9 7.8 3.8 |
Samples
were diluted with zero calibrator.
E.
Time Delay
As
shown
hereafter, assay results remain accurate even when a sample is dispensed 30
minutes after the calibrator has been added to coated tubes.
|
TIME
DELAY |
|||
|
|
0' |
30' |
|
|
Sample
1 (mIU/ml) Sample
2 (mIU/ml) Sample
3 (mIU/ml) |
2.2 4.8 53.6 |
2.3 5.4 54.6 |
|
F.
Hook effect
A
sample spiked with LH up to 1700 mIU/ml gives a signal above the highest
calibrator concentration.
- If the results
obtained for Control 1 and/or Control 2 are not within the range specified on
the vial label, the results cannot be used unless a satisfactory explanation for
the discrepancy has been given.
-
If desirable, each laboratory can make its own pools of control samples,
which should be kept frozen in aliquots.
-
Acceptance criteria for the difference between the duplicate results of
the samples should rely on Good Laboratory Practises
14. Reference intervals
The
values provided below are given only for guidance; each laboratory should
establish its own normal range of values.
The
range is expressed as 2.5% to 97.5% percentiles.
|
Identification |
Number
of subjects |
Mean (mIU/ml) |
Range (mIU/ml) |
|
Children Adult
males Women .
Ovulatory cycles -
Follicular phase (day –12 to -6) -
Ovulatory peak (day 0) -
Luteal phase (day +6 to +12) .
Postmenopausal |
27 69 34 49 63 53 |
2.3 2.7 4.3 19.6 3.3 31.2 |
0.1
– 8.9 1.0
– 5.3 0.8
– 10.4 2.9
– 41.1 0.5
– 7.6 14.4
– 52.8 |
15. Precautions and warning
Safety
For
in vitro diagnostic use only.
This
radioactive product can be transferred to and used only by authorized persons;
purchase, storage, use and exchange of radioactive products are subject to the
legislation of the end user's country. In
no case the product must be administered to humans or animals.
All
radioactive handling should be executed in a designated area, away from regular
passage. A logbook for receipt and
storage of radioactive materials must be kept in the lab.
Laboratory equipment and glassware, which could be contaminated with
radioactive substances, should be segregated to prevent cross contamination of
different radioisotopes.
Any
radioactive spills must be cleaned immediately in accordance with the radiation
safety procedures. The radioactive
waste must be disposed of following the local regulations and guidelines of the
authorities holding jurisdiction over the laboratory. Adherence to the basic rules of radiation safety provides
adequate protection.
The
human blood components included in this kit have been tested by European
approved and/or FDA approved methods and found negative for HBsAg, anti-HCV,
anti-HIV-1 and 2. No known method
can offer complete assurance that human blood derivatives will not transmit
hepatitis, AIDS or other infections. Therefore,
handling of reagents, serum or plasma specimens should be in accordance with
local safety procedures.
All animal
products and derivatives have been collected from healthy animals. Bovine
components originate from countries where BSE has not been reported.
Nevertheless, components containing animal substances should be treated as
potentially infectious.
Avoid
any skin contact with reagents (sodium azide as preservative).
Azide in this kit may react with lead and copper in the plumbing and in
this way form highly explosive metal azides.
During the washing step, flush the drain with a large amount of water to
prevent azide build-up.
Do not smoke,
drink, eat or apply cosmetics in the working area.
Do not pipette by mouth. Use
protective clothing and disposable gloves.
16.
Bibliography
1.
CLARKE J.J.,
CUMMINS J.T. (1987)
Pulsatility
of reproductive hormones: physiological basis and clinical implications.
Bailličre's
Clin. Endocrinol. Metab.,
1:1
2.
FILICORI M.,
SANTORO N., MERRIAN G.R., CROWLEY W.F. Jr., (1986)
Characterization
of the physiological pattern of episodic gonadotropin secretion throughout the
human menstrual cycle.
J.
Clin. Endocrinol. Metab., 62:1136
3.
MAIS V., CETEL N.S., MUSE K.N., QUIGLEY M.E., REID R.L., YEN S.S.C.
(1987)
Hormonal
dynamics during luteal-follicular transition.
J.
Clin. Endocrinol. Metab., 64:1109
4.
SAWYER-STEFFAN J.E., LASLEY B.L., HOFF J.D., YEN S.S.C. (1982)
Comparison
of in-vitro bioactivity and immunoreactivity of serum LH in normal cyclic and
hypogonadal women treated with low doses of LH-RH.
J.
Reprod. Fertil., 65:45
5.
THOMAS C.M.G., SEGERS M.F.G. (1988)
Measurement
of Serum hLH: hCG interference evaluated for two hLH-Specific IRMA kits.
Clin.
Chem., 34:768
6.
LOUMAYE E., VANKRIEKEN L., DEPREESTER S., PSALTI I., de COOMAN S., THOMAS
K. (1989)
Hormonal
changes induced by short-term administration of a gonadotropin-releasing hormone
agonist during ovarian hyperstimulation for in vitro fertilization and their
consequences for embryo development.
Fertil.
and Steril., 51:105
7.
VERMES I., BONTE H.A., SLUIS VEER G., SCHOEMAKER J. (1991)
Interpretations
of five monoclonal immunoassays of Lutropin and Follitropin: effects of
normalization with WHO standard.
Clin.
Chem., 37:415
8.
DEMOULIN A., DUBOIS M., GERDAY C., GILLAIN D., LAMBOTTE R., FRANCHIMONT
P. (1991)
Variation
of luteinizing hormone serum concentration after exogeneous human chorionic
gonadotropin administration during ovarian stimulation.
Fertil.
and Steril., 55:796
9.
MASSA G., de ZECHER F., VANDERSCHUREN-LODEWYK K. (1992)
Serum
levels of immunoreactive Inhibine, FSH and LH in human infants at Preterm and
Term Birth.
Biol.
of the Neonat., 61:150
10.
DE HERTOGH R., VANKRIEKEN L., THOMAS K. de GASPARO M. (1992)
Circhoral
fluctuations of serum total renin, inhibin and related hormones around the
mid-cycle in normal human females.
Hum.
Reprod., 7:337
17.
Summary of the protocol
|
Total Counts (µl) |
Standards (µl) |
Sample(s) (µl) |
|
|
Standards (0-6) Samples Tracer |
- - 50 |
100 - 50 |
- 100 50 |
|
Incubation |
1 hours at RT + 700 rpm |
||
|
Separation Washing solution Separation Washing solution Separation |
- - - - - |
Aspirate (or decant) 2.0 Aspirate (or decant) 2.0 Aspirate (or decant) |
|
|
Counting |
Count tubes for 60 seconds |
||
|