|
|
|
Free
βhCG-IRMA
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Reagents | 100 tests kit | colour code | reconstitution |
| Coated tubes with anti-b-hCG (monoclonal antibodies) | 2x50
|
ready to use
|
|
| Tracer: 125Iodine labelled anti-b-hCG in Phosphate buffer with BSA, azide (<0.1%) and inert red dye | 1vial 22 ml 750kBq | red | ready to use |
| standard zero in human serum with thymol | 1 vial lyoph. | yellow | add 0.5 ml distilled water |
| standard 1 to 7 (see exact values on labels) in human serum with thymol | 5 vials lyoph. | yellow | add 0.5 ml distilled water |
| specimen diluent: Phosphate buffer with BSA and azide | 1 vial 22 ml | black | ready to use |
| incubation buffer: phosphate buffer with BSA and azide | 1 vial 22 ml | black | ready to use |
| wash solution conc. (tris-HCl) | 2 vials 10 ml | brown | dilute 70x with distilled water (use a magnetic stirrer) |
| control 1&2 in human serum with thymol | 2 vials lyoph. | silver | add 0.5 ml distilled water |
Note:
1. Use the specimen
diluent for sera dilutions.
2.
1 mIU
of the calibrator preparation is equivalent to 1 mIU of the 3rd IRP
75/551.
4. Supplies not provided
The
following material is required but not provided in the kit:
1.
Distilled water
2.
Pipettes for delivery of: 50 μl, 200 μl, 500 µl and 1 ml. (the
use of accurate pipettes with disposable plastic tips is recommended)
3.
Vortex mixer
4.
Magnetic stirrer
5.
Tube shaker (600 rpm)
6.
5 ml automatic syringe (Cornwall type) for washing
7.
Aspiration system (optional).
8.
Any gamma counter capable of measuring 125I may be used
(minimal yield 70%).
5. reagents preparation
A.
Calibrators: Reconstitute the calibrators with 0.5 ml distilled water..
B.
Controls: Reconstitute the controls with 0.5 ml distilled water.
C.
Working Wash solution: Prepare an adequate volume of Working Wash
solution by adding 69 volumes of distilled water to 1 volume of Wash Solution
(70x). Use a magnetic stirrer to homogenize. Discard unused Working Wash
solution at the end of the day.
6. Storage and expiration dating of reagents
-
Before opening or reconstitution, all kit components are stable until the
expiry date, indicated on the label, if kept at 2 to 8°C.
-
After reconstitution, calibrators and controls are stable for 7 days at
2-8°C.
For longer storage periods, aliquots should be made and kept at –20°C
for 3 months. Avoid subsequent
freeze-thaw cycles.
-
Freshly prepared Working Wash solution should be used on the same day.
- After its first use, tracer
is stable until expiry date, if kept in the original well-closed vial at 2 to 8°C.
- Alterations in physical
appearance of kit reagents may indicate instability or deterioration.
7. Specimens collection and preparation
Serum must be kept at 2 – 8°C.
If the test is not run within 24 hours, storage at –20°C
is recommended.
Avoid subsequent freeze-thaw cycles.
8. Procedure
A.
Handling notes
Do
not use the kit or components beyond expiry date. Do not mix materials from different kit lots.
Bring
all the reagents to room temperature prior to use.
Thoroughly
mix all reagents and samples by gentle agitation or swirling. In order to avoid
cross-contamination, use a clean disposable pipette tip for the addition of each
reagent and sample.
High
precision pipettes or automated pipetting equipment will improve the precision.
Respect the incubation times.
Prepare a calibration
curve for each run, do not use data from previous runs.
B.
Procedure
1.
Label
coated tubes in duplicate for each calibrator, control and sample.
For determination of total counts, label 2 normal tubes.
2.
Briefly
vortex
calibrators, samples and controls and dispense 50 μl
of each into the respective tubes.
3.
Dispense 200 µl of Incubation Buffer into each tube except total counts.
4.
Incubate
for 30 min at room temperature
with continuous shaking (600 rpm).
5.
Aspirate
(or decant) the content of each tube (except total counts).
Be sure that the plastic tip of the aspirator reaches the bottom of the
coated tube in order to remove all the liquid.
6.
Wash
the tubes twice with 3 ml Wash Solution (except total counts).
Avoid foaming during the addition of the Working Wash Solution.
7.
Aspirate
(or decant) the content of each tube (except total counts).
8.
After the last washing, let the
tubes standing upright for two minutes and aspirate the remaining drop of
liquid.
9.
Dispense
200 μl
of anti-βhCG-125I tracer into each tube, including the uncoated
tubes for total counts.
10.
Shake the rack containing the tubes gently by hand to liberate any
trapped air bubbles.
11.
Incubate
for 30 min at room temperature
with continuous shaking (600 rpm).
12.
Aspirate
(or decant) the content of each tube (except total counts).
Be sure that the plastic tip of the aspirator reaches the bottom of the
coated tube in order to remove all the liquid.
13.
Wash
the tubes with 3 ml Wash Solution (except total counts).
Avoid foaming during the addition of the Working Wash Solution.
14.
Aspirate
(or decant) the content of each tube (except total counts).
15.
Let the
tubes standing upright for two minutes and aspirate the remaining drop of
liquid.
16.
Count
the tubes in a gamma counter for 60 seconds.
9.
Calculation of results
1.
Calculate the mean of duplicate determinations.
2.
On semi-logarithmic or linear graph paper plot the c.p.m. (ordinate) for
each calibrator against the corresponding concentration of βhCG
(abscissa) and draw a calibration curve through the calibrator points, reject
the obvious outliers.
3.
Read the concentration for each control and sample by interpolation on
the calibration curve.
4.
Computer assisted data reduction will simplify these calculations. If automatic result processing is used, a 4-parameter
logistic function curve fitting is recommended.
10. Typical data
The
following data are for illustration only and should never be used instead of the
real time calibration curve.
|
|
|
|
|
|
Total
count |
230469 |
|
|
|
Calibrator |
0 mIU/ml
0.29
mIU/ml
0.96
mIU/ml
2.8
mIU/ml
8.8
mIU/ml
28
mIU/ml
56
mIU/ml
93
mIU/ml |
299
1196
3084
8790
25501
68274 108661
146141 |
0.1 0.5 1.3 3.8 11.1 29.6 47.1 63.4 |
11. Performance and limitations
A.
Detection Limit
Twenty
zero calibrators were assayed along with a set of other calibrators. The
detection limit, defined as the apparent concentration two standard deviations
above the average counts at zero binding, was 0.03 mIU/ml.
B.
Specificity
Cross-reactive
hormones were added to a low and to a high βhCG value calibrator.
The apparent
βhCG
response was measured.
|
added
Hormone |
βhCG
CAL 1 mIU/ml |
βhCG
CAL 6 mIU/ml |
|
- |
0.29 |
56 |
|
FSH
300 mIU/ml |
0.28 |
58 |
|
TSH
300 µIU/ml |
0.25 |
58 |
|
LH
300 mIU/ml |
0.31 |
60 |
|
βLH
10000 ng/ml |
46.8 |
86 |
|
αhCG
500 ng/ml |
0.8 |
55 |
C.
Precision
|
INTRA
ASSAY |
|
||||||
|
Serum |
N |
<X>
±
S.D. (mIU/ml) |
CV (%) |
Serum |
N |
<X>
±
S.D. (mIU/ml) |
CV (%) |
|
A B |
10 10 |
1.42
± 0.10 10.7
± 0.3 |
7.3 2.5 |
C D |
20 20 |
1,32
± 0.08 8,9
± 0,3 |
5,8 3, |
D.
Accuracy
RECOVERY TEST
|
Sample |
Added
βhCG (mIU/ml) |
Recovered
βhCG (mIU/ml) |
Recovery (%) |
|
A |
0.44 0.81 3.66 19.5 37.3 |
0.38 0.79 3.88 19.0 36.7 |
87.2 97.9 106.0 97.4 98.4 |
|
DILUTION TEST |
|||
|
Sample
|
Dilution |
Theoretical
Concent. (mIU/ml) |
Measured
Concent. (mIU/ml) |
|
1 2 |
1/1 1/2 1/5 1/10 1/20 1/1 1/2 1/5 1/10 1/20 |
- 3.87 1.55 0.77 0.39 - 15.31 6.12 3.06 1.53
|
7.73 4.21 1.31 0.76 0.43 30.62 15.29 6.13 2.88 1.51 |
Samples
were diluted with specimen diluent.
E.
Time Delay
As
shown below, assay results remain accurate even when a sample is dispensed up to
30 minutes after the calibrator has been added to the coated tubes.
|
TIME
DELAY |
||||
|
|
0'
(mIU/ml) |
10'
(mIU/ml) |
20'
(mIU/ml) |
30'
(mIU/ml) |
|
Serum
1 Serum
2 |
1.33 10.27 |
1.35 9.96 |
1.34 10.56 |
1.35 10.49 |
F.
Hook effect
A
serum sample with a concentration of 12800 mIU/ml βhCG gives a signal above
the highest calibrator concentration.
12. Internal Quality control
- If the results
obtained for Control 1 and/or Control 2 are not within the range specified on
the vial label, the results cannot be used unless a satisfactory explanation for
the discrepancy has been given.
-
If desirable, each laboratory can make its own pools of control samples,
which should be kept frozen in aliquots.
-
Acceptance criteria for the difference between the duplicate results of
the samples should rely on Good Laboratory Practises
13. Reference intervals
The
values provided below are given only for guidance; each laboratory should
establish its own normal range of values.
|
Identification |
Number
of subjects |
Values (mIU/ml) |
|
Normal
subjects |
33 |
<
0.1 mIU/ml |
14. Precautions and warning
For
in vitro diagnostic use only.
This
radioactive product can be transferred to and used only by authorized persons;
purchase, storage, use and exchange of radioactive products are subject to the
legislation of the end user's country. In
no case the product must be administered to humans or animals.
All
radioactive handling should be executed in a designated area, away from regular
passage. A logbook for receipt and
storage of radioactive materials must be kept in the lab.
Laboratory equipment and glassware, which could be contaminated with
radioactive substances, should be segregated to prevent cross contamination of
different radioisotopes.
Any
radioactive spills must be cleaned immediately in accordance with the radiation
safety procedures. The radioactive
waste must be disposed of following the local regulations and guidelines of the
authorities holding jurisdiction over the laboratory. Adherence to the basic rules of radiation safety provides
adequate protection.
The
human blood components included in this kit have been tested by European
approved and/or FDA approved methods and found negative for HBsAg, anti-HCV,
anti-HIV-1 and 2. No known method
can offer complete assurance that human blood derivatives will not transmit
hepatitis, AIDS or other infections. Therefore,
handling of reagents, serum or plasma specimens should be in accordance with
local safety procedures.
All
animal products and derivatives have been collected from healthy animals. Bovine
components originate from countries where BSE has not been reported.
Nevertheless, components containing animal substances should be treated as
potentially infectious.
Avoid
any skin contact with reagents (sodium azide as preservative).
Azide in this kit may react with lead and copper in the plumbing and in
this way form highly explosive metal azides.
During the washing step, flush the drain with a large amount of water to
prevent azide build-up.
Do
not smoke, drink, eat or apply cosmetics in the working area.
Do not pipette by mouth. Use
protective clothing and disposable gloves.
15.
Bibliography
1.
CHEN F. et al., (1987)
Radioimmunoassay
of the serum free β-subunit of human chorionic gonadotropin in
trophoblastic disease.
J.
Clin. Endocrinol. Metab., 64:313.
2.
DAWOOD M.Y. et al., (1977)
Human
chorionic gonadotropin and its subunit in hydatidiform mole and choriocarcinoma.
Obstet.
Gynecol., 50:172.
3.
GASPARD V.J. et al., (1980)
Serum
concentration of human chorionic gonadotropin and its alpha and beta subunit in
Trophoblastic tumours.
Clin.
Endocrinol. (Oxf.), 13:219.
4.
PIERCE J.G. et al., (1981)
Glycoprotein
hormones : structure and function.
Annu.
Rev. Biochem., 50:465.
5.
KARDANO A. et al., (1994)
Human
Chorionic Gonadotropin β-Subunit Nicking enzymes in Pregnancy and Cancer
Patient Serum.
J.
Clin. Endocr. Metab., 79:761-767.
16. Summary of the protocol
|
Total Counts (µl) |
Standards (µl) |
Sample(s) (µl) |
|
|
Standards (0-7) Samples incubation buffer |
- - 200 |
50 - 200 |
- 50 200 |
|
Incubation |
30 min at RT + continuos shaking |
||
|
Separation Washing solution Separation Washing solution Separation |
- - - - - |
Aspirate (or decant) 3.0 Aspirate (or decant) 3.0 Aspirate (or decant) |
|
|
Tracer |
200 |
200 |
200 |
|
Incubation |
30 min at RT + continuos shaking |
||
|
Separation Washing solution Separation |
- - - |
Aspirate (or decant) 3.0 Aspirate (or decant) |
|
|
Counting |
Count tubes for 60 seconds |
||
|