|
|
|
hCG
+ β - Elisa
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Reagents |
Quantity |
Colour
code |
Reconstitution |
|
Microtiter
plate with 96 anti-hCG coated wells |
96
wells |
blue |
Ready
for use |
|
Standard
0 mIU/ml in bovine serum with preservatives |
1
vial 10
ml |
yellow |
Ready
for use |
|
Standards
1-5 in bovine serum with preservatives (3rd IS 75/537) (see vial label for exact concentrations) |
5
vials lyophilized |
yellow |
Add
0.5 ml distilled water |
|
Control
1 and 2 in human plasma with preservatives |
2
vials lyophilized |
silver |
Add
0.5 ml distilled water |
|
Anti-hCG-HRP
conjugate in Tris buffer with bovine serum albumin, maleate and
preservatives |
1
vial 6
ml |
red |
Ready
for use |
|
Washing
Solution concentrate (buffer with preservatives) |
1
vial 10
ml |
brown |
Dilute
2 ml in 400 ml distilled water or the vial content in 2000 ml distilled
water (use a magnetic stirrer) |
|
Chromogenic
Solution : TMB (Tetramethylbenzidine) |
1
vial 25
ml |
brown |
Ready
for use |
|
Stop
Solution |
1
vial 25
ml |
white |
Ready
for use |
Note
: Use the content of the zero standard for
dilution of samples.
IV.
PRECAUTIONS
AND WARNINGS
1.
The human blood components included in this kit have been tested by
European approved and USA FDA approved methods and found negative for HBsAg,
anti-HCV and anti-HIV-1 and 2. No known method can offer complete assurance that
human blood derivatives will not transmit hepatitis, AIDS or other infections.
Therefore, handling of reagents, serum, or plasma specimens should be in
accordance with local safety procedures.
2.
Avoid any skin contact with Stop Solution and Chromogenic Solution (TMB). In case of contact wash thoroughly with water.
3.
Do not eat, drink, smoke or apply cosmetics where kit reagents are used.
4. Do not pipette liquids by
mouth.
V.
SUPPLIES NOT PROVIDED
1. High quality distilled water.
2. Precision pipette : 50 μl,
100 μl, 0.5 ml and 2 ml.
3.
Vortex mixer and magnetic stirrer.
4.
Horizontal microtiter plate
shaker capable of 700 rpm ± 100 rpm, microtiter plate reader capable of reading
at 450 nm and at any wavelength from 600 to 750 nm, microtiter plate washer.
VI
REAGENTS
PREPARATION
1. Standards and Controls :
Reconstitute the lyophilized Standards and controls to the volume specified on
the vial label with distilled water. Allow
the vials to remain undisturbed until completely dissolved, then mix well by
gentle inversion.
2.
Wash working solution : dilute 2 ml in 400 ml distilled water or the
vial content in 2000 ml distilled water (use a magnetic stirrer)
VII
STORAGE
AND
SHELF
LIFE
OF
REAGENTS
A. Unopened vials
Store
the unopened kit at 2°C to 8°C. All kit components are
stable until the expiry date printed on the labels.
B. Opened vials
1.
The Conjugate vial must be stored at 2°C to 8°C.
2.
The reconstituted Standards and Controls are stable for 3 days at 2°C to
8°C. Aliquots held for longer periods of time should be frozen, a maximum of
two times, at -20°C (maximum 2 months) or at -70°C for longer storage (until
expiration date).
3.
Store the unused strips at 2°C to 8°C in the sealed bag containing the
desiccant until expiration date.
4.
The Wash Solution Concentrate is stable at room temperature until
expiration date. In order to avoid washer obstructions, it is recommended to
prepare a fresh diluted Wash Solution each day.
VII
SPECIMEN
COLLECTION,
PREPARATION,
STORAGE
AND DILUTION
A. Specimen collection and preparation
-
Prior to use, all the samples should be at room temperature.
It’s recommended to vortex the samples before use.
Hemolysis has to be avoided.
-
Serum, heparinized plasma or EDTA plasma provide similar results and may
be used.
Y(serum)
= 0.97 x (EDTA plasma) + 0.21 R =
0.99 n = 34
Y(serum)
= 0.96 x (hepar. plasma) + 0.79 R =
0.99 n = 34
B. Storage
Serum/plasma
samples must be kept at 2°C to 8°C for maximum
hours, and for longer storage (maximum one year) at ‑20°C.
C. Sample dilution
If
a specimen is expected or known to have a concentration above the highest
standard it has to be diluted with the zero standard to fall within the
measuring interval.
IX.
PROCEDURE
The
instructions of the assay procedure must be followed to obtain reliable results.
A.
Procedural notes
1.
Allow the samples and reagents to equilibrate to room temperature (18°C
to 25°C) before commencing the assay. Thoroughly mix the reagents and samples
before use by gentle agitation or swirling.
2.
Do not use kit components beyond the expiration date.
3.
Do not mix materials from different kit lots.
4.
Do not mix strips from different plates.
5.
Perform Standards,Controls and Unknowns in duplicate. Vertical alignment
is recommended.
6.
A standard curve should be run with each assay run or each plate run.
7.
To avoid drift, the time between pipetting of the first standard and the
last sample must be no longer than 30 minutes.
Use
a clean disposable pipette tip for the addition of each reagent, standard,
control or specimen in order to avoid cross contamination.
9.
For the dispensing of the Chromogenic Solution and Stop Solution avoid
pipettes with metal parts.
10.
Use a clean plastic container to prepare the Wash Solution.
11.
The Chromogenic Solution should be colourless. If a blue colour develops
before use, this indicates that the reagent is unusable, and must be discarded.
12.
During incubation with Chromogenic Solution, avoid direct sunlight on the
microtiter plate.
13. Respect
the incubation times described in the assay procedure.
B. Assay procedure
1.
Select the required number of strips for the run. The unused strips
should be resealed in the bag with desiccant and stored at 2-8°C.
2.
Secure the strips into the holding frame.
3.
Pipette 50 μl of each Standards, Controls or Samples into the
appropriate wells. Time between
distribution of first standard and last sample can be up to 30 minutes without
affecting the results. Vertical
alignment is recommended
4.
Pipette 50 μl of conjugate into each well.
5.
Incubate for 30 minutes at room temperature (20°C - 27°C) on a
horizontal shaker set at 700 rpm ± 100 rpm.
6.
Aspirate the liquid from each well ;
7.
Wash the plate three times by :
a)
dispensing of 0.3 ml of BioSource Wash Solution into each well ;
b)
aspirating the content of each well.
8.
Pipette 100 μl of the Chromogenic Solution into each well
within 15 min. following the washing step.
9.
Incubate the plate for 15 min. at room temperature on an horizontal
shaker set at 700 ± 100 rpm,
avoiding direct sunlight.
10.
Pipette 100 μl of Stop Solution into each well.
11.
Read absorbances at 450 nm (reference filter : 650 nm) within 1 hour
and calculate the results as described in section XI.
X
CALCULATION
OF ANALYTICAL
RESULTS
. Read the microtiter
plate at 450 nm (reference filter : 650 nm).
.
Plot the OD on the ordinate against the standard concentrations on the
abscissa using either linear or semi-log graph paper and draw the curve by
connecting the plotted points with straight lines.
.
Determine the hCG + β concentration of the samples from the standard
curve.
Example
of a typical standard curve
The
following data are for demonstration purpose only and can not be used in place
of data generated at the time of assay.
|
hCG
+ β - ELISA (mIU/ml) |
mOD
units (450
nm/650 nm) |
|
Standard
0 5 15 50 150 500 |
50 120 275 855 2262 2955 |
XI
QUALITY
CONTROL
-
The two Controls provided in the kit can be used as internal laboratory
controls.
Note
: Other controls which contain azide will
interfere with the enzymatic reaction and cannot be used.
-
Serum or heparin plasma pools can be collected and frozen immediately in
aliquot to serve as controls. Repeated freezing and thawing are not permitted.
-
Record keeping : it is good laboratory practice to record the kit
lot numbers and date of reconstitution for the reagents in use.
-
Controls : it is recommended that Controls be routinely assayed as
unknown samples to measure assay variability. It is recommended that quality
controls charts be maintained to monitor the performance of the kits. Control
ranges are indicated on vial labels. Out of range control results indicate the
assay must be repeated. Repeat patient samples may also be used to measure
interassay precision.
-
Sample handling : strictly adhere to the instruction for handling
and storage of samples. Standards, Controls, and Unknowns should be run in
duplicate. A clean disposable tip should always be used to avoid carryover
contamination.
-
Data reduction : it is good practice to construct a standard curve
for each run to check visually the curve fit selected by the computer program.
XII
EXPECTED VALUES
These
are provided for guidance only. Each
lab should establish its own normal range.
|
Healthy
adults |
Range (mIU/ml) |
|
|
|
|
Males
(n = 244) Pre-menopausal
women (n = 195) Post-menopausal
women (n = 50) Pregnant
women : 1st
week 2nd
week 3rd
week 4th
week 5th-9th
week 10th-13th
week 2nd
quarter 3rd
quarter |
<
0.6 - 4 <
0.6 - 5 <
0.6 - 4 up
to 50 40
- 4000 50
- 15000 400
- 43000 2500
- 220000 15000
- 230000 4000
- 75000 1000
- 65000 |
XIII
PERFORMANCE CHARACTERISTICS
1. Minimum Detectable Concentration
(MDC).
The
MDC is estimated to be 0.6 mIU/ml and is defined as the hCG + β
concentration corresponding to the average OD of 20 replicates of the zero
standard - 2 standard deviations.
2. Precision
|
INTRA-ASSAY |
INTER-ASSAY |
||||||
|
Serum |
n |
<X>
± SD (pmol/ml) |
CV (%) |
Serum |
n |
<X>
± SD (pmol/ml) |
CV (%) |
|
1 2 |
28 28 |
27.7
± 1.0 130.7
± 4.0 |
3.7 2.9 |
1 2 |
10 10 |
50.9
± 2.1 99.6
± 5.0 |
4.1 5.1 |
3. Specificity
The
cross-reactivity of FSH, LH, TSH, αhCG, βhCG was determined by
addition of each analyte to serum samples respectively containing 36.6 and 156.4
mIU/ml of hCG and measuring the apparent hCG concentration.
As shown below, the cross-reactions with hLH, hFSH, TSH and αhCG are
insignificant while βhCG totally cross-reacts with hCG.
|
Added
analyte to a serum containing 36.5 mIU hCG/ml (=112
fmol/ml) (*) |
Observed
hCG value (mIU/ml) |
Added
analyte to a serum containing 156.4 mIU hCG/ml (=485
fmol/ml) (*) |
Observed
hCG value (mIU/ml) |
||
|
FSH LH TSH αhCG βhCG |
200
mIU/ml 200
mIU/ml 200
mIU/ml 1000
ng/ml 1100
fmol/ml |
36.6 34.5 35.5 35.7 340 =1054
fmol/ml |
FSH LH TSH αhCG βhCG |
200
mIU/ml 200
mIU/ml 200
mIU/ml 1000
ng/ml 1100
fmol/ml |
158.4 154.2 156.8 159.1 459 =1422
fmol/ml |
(*)
Conversion assuming a MW for hCG of 37900 and 8.6 mIU/ng (3rd IRP
75/537)
4. Accuracy
|
RECOVERY |
DILUTION
TEST |
|
Sam-ple |
Added hCG (mIU/ml) |
Recov. hCG (mIU/ml) |
Reco very (%) |
Serum dilu-tion |
Theor. Conc. (mIU/ml) |
Meas. Conc. (mIU/ml) |
Reco- very (%) |
|
Serum Plasma |
6.4 24.0 65.0 247.6 6.4 24.0 65.0 247.6 |
6.1 23.5 58.4 230.2 6.7 23.9 63.9 248.9 |
95 98 90 93 105 100 98 100 |
1/1 1/2 1/4 1/8 1/16 1/32 1/4 1/8 1/16 1/32 1/64 |
320 160 80 40 20 10 177 88.5 44.2 21.1 11.1 |
- 162 90 45 20.2 10.0 - 85.8 42.7 19.3 10.2 |
- 101 112 112 101 100 - 97 97 92 92 |
5. High dose hook-effect
A
sample spicked with hCG up to 250 000 mIU/ml gives a result higher than the last
standard point.
XIV
LITERATURE REFERENCES
1.
BRAUNSTEIN G.D., RASOR J., ADLER D., DANZER H., WADE H.E., (1976)
Serum
human vhorionic gonadotropin levels throughout normal pregnancy.
Am.
J. Obstet. Gynecol., 126:678.
2.
CHIEN F., SETSUKO G., YOSHIHITO F., YUTAKA T., (1987)
Radioimmunoassay
of the serum free beta-subunit of human chorionic gonadotropin in trophoblastic
disease.
J.
Clin. Endocrinol. Metab., 64:313.
3.
DAWOOD M.Y., SASCENA B.B., LANDESMAN R., (1977)
Human
chorionic gonadotropin and its subunit in hydatidiform mole and choriocarcinoma.
Obstet.
Gynecol., 50:172.
4.
HAY D.L., (1982)
Chorionic
gonadotropin.
Clin.
Biochem. Rev., 3:35.
5.
PIERCE J.B., PARSONS T.F., (1981)
Glycoprotein
hormones : structure and function.
Annu.
Rev. Biocham., 50:465.
XV
SUMMARY OF ASSAY PROCEDURE
|
|
Standards (μl) |
Control-Samples (μl) |
|
Standards
(0-5) Controls-samples Anti
hCG-HRP |
50 - 50 |
- 50 50 |
|
Incubate
for 30 min at RT (20°C - 27°C) with continuous shaking (700 RPM) Aspirate
the content of each well Wash
3 times with 0.4 ml of wash solution and aspirate |
||
|
Chromogenic
Solution |
100 |
100 |
|
Incubate
for 15 min. at RT with continuous shaking (700 RPM) |
||
|
Stop
solution |
100 |
100 |
|
Read
the microtiter plate at 450 nm (versus 650 nm) |
||
|