TSH-Elisa

Immunoenzymetric assay for the in vitro quantitative measurement of human Thyroid Stimulating
Hormone (TSH) in serum and plasma.
BL-13-E
For in
vitro diagnostic use only
I.
CLINICAL BACKGROUND
Biological activities
Thyrotrope
cells of the anterior pituitary synthesize and secrete human thyroid
stimulating hormone (TSH), a glycoprotein of molecular weight 28,000 Da,
comprising two subunits : α-TSH is very similar to a α subunit of
LH, FSH and hCG, β-TSH differs from other hormone subunits and defines
the immunological specificity.
TSH
regulates the synthesis and release of thyroid hormones : thyroxin (T4) and
triiodothyronine (T3). TSH
secretion is stimulated by a hypothalamic peptide, TRH (TSH releasing
hormone); a negative feedback on TSH secretion is exerted by T3 and T4.
II.
PRINCIPLES OF
THE METHOD
The
Bio-Line TSH-ELISA is a solid phase Enzyme Amplified Sensitivity Immunoassay
performed on microtiterplates.
Calibrators and samples react with the capture monoclonal antibody (MAb
1) coated on microtiter well and with a monoclonal antibody (MAb 2) labelled
with horseradish peroxidase (HRP). After
an incubation period allowing the formation of a sandwich: coated MAb 1 –
human TSH – MAb 2 – HRP, the microtiterplate is washed to remove unbound
enzyme labelled antibody. Bound
enzyme-labelled antibody is measured through a chromogenic reaction.
Chromogenic solution (TMB – H2O2) is added and
incubated. The reaction is stopped
with the addition of Stop Solution and the microtiterplate is then read at the
appropriate wavelength. The
amount of substrate turnover is determined colourimetrically by measuring the
absorbance, which is proportional to the TSH concentration.
A
calibration curve is plotted and TSH concentration in samples is determined by
interpolation from the calibration curve. The
use of the ELISA reader (linearity up to 3 OD units) and a sophisticated data
reduction method (polychromatic data reduction) result in a high sensitivity in
the low range and in an extended calibration range.
III.
REAGENTS PROVIDED
|
Reagents
|
96
tests
Kit
|
Color
Code
|
Reconstitution
|
|
Microtiterplate
with 96 anti TSH (monoclonal antibodies) coated wells
|
96
wells
|
blue
|
Ready
for use
|
|
Conjugate:
HRP labelled anti-TSH (monoclonal antibodies) in TRIS-HCl buffer with
bovine serum albumin and thymol
|
1
vial
6
ml
|
red
|
Ready
for use
|
|
Zero
calibrator in horse serum
and thymol
|
1
vial
lyophilized
|
yellow
|
Add
2.0 ml distilled water
|
|
Calibrator
N = 1 to 6
(see
exact values on vial labels) in horse serum and thymol
|
6
vials
lyophilized
|
yellow
|
Add
2.0 ml distilled water
|
|
Wash
Solution (Tris-HCl)
|
1
vial
10
ml
|
brown
|
Dilute
200 x with distilled water (use a magnetic stirrer).
|
|
Controls
- N = 1 or 2
in
human serum with thymol
|
2
vials
lyophilized
|
silver
|
Add
1.0 ml distilled water
|
|
Chromogen
TMB (Tetramethylbenzydine) in Dimethylformamide
|
1
vial
1
ml
|
green
|
Dilute
0.2
ml into 1 vial of substrate buffer
|
|
Substrate
buffer: H2O2 in acetate / citrate buffer
|
3
vials
21
ml
|
white
|
Ready
for use
|
|
Stop
solution: H2SO4 1.8N
|
1
vial
6
ml
|
black
|
Ready
for use
|
Note:
1.
Use the zero calibrator for sample dilutions.
2. 1
µIU of the calibrator is equivalent to 1 µIU of the 2nd IRP 80/558.
IV.
SUPPLIES NOT PROVIDED
The
following material is required but not provided in the kit:
1.
High quality distilled water
2.
Pipettes for delivery of: 50 μl, 100 μl, 200 µl, 1 ml and 2 ml
(the use of accurate pipettes with disposable plastic tips is recommended)
3.
Vortex mixer
4.
Magnetic stirrer
5.
Horizontal microtiterplate shaker capable of 700 rpm ± 100 rpm
6.
Washer for microtiterplates
7.
Microtiterplate reader capable of reading at 450 nm, 490 nm and 650 nm
(in case of polychromatic reading) or capable of reading at 450 nm and 650 nm
(monochromatic reading)
8.
Optional equipment: The ELISA-AID™ necessary to read the plate
according to polychromatic reading (see paragraph XI.A.) can be purchased from
Robert Maciels Associates, Inc. Mass. 0.2174 USA.
V.
REAGENT PREPARATION
a.
Calibrators
: Reconstitute the calibrators with 2.0 ml distilled water.
b.
Controls
: Reconstitute the controls with 1.0 ml distilled water.
c.
Working Wash solution
: Prepare an adequate volume of Working Wash solution by adding 199 volumes of
distilled water to 1 volume of Wash Solution (200x). Use a magnetic stirrer to
homogenize. Discard unused Working Wash solution at the end of the day.
d.
Revelation Solution:
pipette 0.2 ml of the chromogen TMB into one of the vials of substrate buffer (H2O2
in acetate/citrate buffer). Extemporaneous
preparation is recommended.
VI.
STORAGE
AND EXPIRATION
DATING OF
REAGENTS
§
Before opening or reconstitution, all kit components are stable until the
expiry date, indicated on the vial label, if kept at 2 to 8°C.
§
Unused strips must be stored, at 2-8°C, in a sealed bag containing a
desiccant until expiration date.
§
After reconstitution, calibrators and controls are stable for 1 week at 2
to 8°C. For longer storage periods, aliquots should be made and kept at
‑20°C. Avoid successive
freeze thaw cycles.
§
The concentrated Wash Solution is stable at room temperature until
expiration date.
§
Freshly prepared Working Wash solution should be used on the same day.
§
After its first use, the conjugate is stable until expiry date, if kept
in the original well-closed vial at 2 to 8°C.
§
The freshly prepared revelation solution is stable, before use, for
maximum 15 minutes at room temperature and must be discarded afterwards.
§
Alterations in physical appearance of kit reagents may indicate
instability or deterioration.
VII.
SPECIMEN
COLLECTION
AND PREPARATION
§
Serum and plasma must be kept at 2 - 8°C.
§
If the test is not run within 24 hours, storage in aliquots at -20°C is
recommended. Avoid subsequent
freeze thaw cycles.
§
Prior to use, all samples should be at room temperature.
It is recommended to vortex the samples before use.
§
Serum, heparinized plasma or EDTA plasma provide similar results.
Y(serum)
= 1.06 x (EDTA plasma) + 0.15 µIU/ml r=0.96
n=53
Y(serum)
= 0.98 x (Heparin plasma) + 0.07 µIU/ml r=1.0
n=53
§
Do not use haemolysed samples.
VIII.
PROCEDURE
A.
Handling notes
Do
not use the kit or components beyond expiry date.
Do
not mix materials from different kit lots.
Bring
all the reagents to room temperature prior to use.
Thoroughly
mix all reagents and samples by gentle agitation or swirling.
Perform
calibrators, controls and samples in duplicate.
Vertical alignment is recommended.
Use
a clean plastic container to prepare the Wash Solution.
In
order to avoid cross-contamination, use a clean disposable pipette tip for the
addition of each reagent and sample.
For
the dispensing of the Revelation Solution and the Stop Solution avoid pipettes
with metal parts.
High
precision pipettes or automated pipetting equipment will improve the precision.
Respect
the incubation times.
To
avoid drift, the time between pipetting of the first calibrator and the last
sample must be limited to the time mentioned in section XIII paragraph E (Time
delay).
Prepare
a calibration
curve for each run, do not use data from previous runs.
The
chromogenic solution should be colourless.
If a dark blue colour develops within a few minutes after preparation,
this indicates that the preparation is unusable and must be discarded.
Dispense
the Revelation Solution within 15 minutes following the washing of the
microtiterplate.
During
incubation with Revelation Solution, avoid direct sunlight on the
microtiterplate.
B.
Procedure
1.
Select the required number of strips for the run.
The unused strips should be resealed in the bag with a desiccant and
stored at 2-8°C.
2.
Secure the strips into the holding frame.
3.
Pipette 100 µl of each Calibrator, Control and Sample into the
appropriate wells.
4.
Pipette 50 µl of anti-TSH-HRP conjugate into all the wells.
5.
Incubate for 1 hour at room temperature on a horizontal shaker set at 700
rpm ± 100 rpm.
6.
Aspirate the liquid from each well.
7.
Wash the plate 3 times by:
§
dispensing 0.4 ml of Wash Solution into each well
§
aspirating the content of each well
8.
Pipette 200 µl of the freshly prepared revelation solution into each
well within 15 minutes following the washing step.
9.
Incubate the microtiterplate for 15 minutes at room temperature on a
horizontal shaker set at 700 rpm ± 100 rpm, avoid direct sunlight.
10.
Pipette 50 µl of Stop Solution into each well.
11.
Read the absorbances at 450 nm and 490 nm (reference filter 630 nm or 650
nm) within 1 hour and calculate the results as described in section XI.
IX.
CALCULATION
OF RESULTS
A.
Polychromatic Reading:
1.
In this case, the ELISA-AID™ software will do the data processing.
2.
The plate is first read at 450 nm against a reference filter set at 650
nm (or 630 nm).
3.
A second reading is performed at 490 nm
against the same reference filter.
4.
The ELISA-AID™ Software will drive the reader automatically and will
integrate both readings into a polychromatic model.
This technique can generate OD’s up to 10.
5.
The principle of polychromatic data processing is as follows:
§
Xi = OD at 450 nm
§
Yi = OD at 490 nm
§
Using a standard unweighted linear regression, the parameters A & B
are calculated : Y = A*X – B
§
If Xi < 3 OD units, then X calculated = Xi
§
If Xi > 3 OD units, then X calculated = (Yi-B)/A
§
A 4-parameter logistic curve fitting is used to build up the calibration
curve.
§
The TSH concentration in samples is determined by interpolation on the
calibration curve.
B.
Bichromatic Reading
1.
Read the plate at 450 nm against a reference filter set at 650 nm (or 630
nm).
2.
Calculate the mean of duplicate determinations.
3.
On semi-logarithmic or linear graph paper plot the OD values (ordinate)
for each calibrator against the corresponding concentration of TSH (abscissa)
and draw a calibration curve through the calibrator points by connecting the
plotted points with straight lines.
4.
Read the concentration for each control and sample by interpolation on
the calibration curve.
5.
Computer assisted data reduction will simplify these calculations.
If automatic result processing is used, a 4-parameter logistic function
curve fitting is recommended.
X.
TYPICAL DATA
The
following data are for illustration only and should never be used instead of the
real time calibration curve.
XIV.
PRECAUTIONS
AND
WARNINGS
Safety
For
in vitro diagnostic use only.
The
human blood components included in this kit have been tested by European
approved and/or FDA approved methods and found negative for HBsAg, anti-HCV,
anti-HIV-1 and 2. No known method
can offer complete assurance that human blood derivatives will not transmit
hepatitis, AIDS or other infections. Therefore,
handling of reagents, serum or plasma specimens should be in accordance with
local safety procedures.
All
animal products and derivatives have been collected from healthy animals. Bovine
components originate from countries where BSE has not been reported.
Nevertheless, components containing animal substances should be treated as
potentially infectious.
Avoid
any skin contact with all reagents, Stop
Solution contains H2SO4, the chromogen contains TMB in
Dimethylformamide, Substrate buffer contains H2O2.
In case of contact, wash thoroughly with water.
Do
not smoke, drink, eat or apply cosmetics in the working area.
Do not pipette by mouth. Use
protective clothing and disposable gloves.
XV.
BIBLIOGRAPHY
1.
BAYER M.F., KRISS J.P., McDOUGALL l.R.
Clinical
experience with sensitive thyrotropin measurements: diagnostic and therapeutic
Implication.
J.
Nucl. Med. 1985, 26:1248‑56.
2.
CALDWELL G., KELLET H.A., GOW S.M., et al.
A
new strategy for thyroid function testing.
Lancet
1985,i: 1117‑9
3.
FIELD J.B.
Pituitary
thyrotropin: Mecanism of action.
In
the thyroid, S. Werner and S.H. Ingbar. Ed. Harper Row, Hagestown, M.D.., 1978.
4.
MARDELL R.T., GAMBER T.R., WINTON M.R. J.
High
sensitivity assay of thyroid stimulating hormone in patients receiving thyroxine
for primary hypothyroidism and thyroid carcinoma.
Br.
Med. J. 1985: 290:335-356.
5.
MUSTO J.D., PIZZOLANTE J.M. and CHESARONE V.P.
A
comment on thyrotropin measurement and evaluation.
Clin.
Chem. 1984;30:329
6.
PIERCE J.C.
Pituitary
thyrotropin: Chemistry.
In
the thyroid., S. Wemer and S.H. Ingbar. Ed. Harper Row, Hagerstown, M.D. 1978.
7.
RODDIS M.J., BURRIN J.M., JOHANNSSEN A., et al.
Serum
thyrotropin: a first-line discriminatory test of thyroid function.
Lancet
1985; i: 277-8
8.
ROSS D.S.
New
Sensitive immunoradiometric assays for thyrotropin (Review).
Ann.
Intern. Med. 1986; 104;718-21
9.
PETER S.A. et al.
Elevated
serum thyrotropin (TSH) levels in critically ill patients with acquired
immunodeficiency syndrome (AIDS).
Exp.
Clin. Endocrinol. 1993; 101(6);346-9
10.
JAIMELA E. et al.
Ability
of two new thyrotropin (TSH) assays to separate hyperthyroid patients from
euthyroid patients with ow TSH.
Clin.
Chem. 1994, Jan;40(1);101-105
11.
KOMOROWSKI J. et al.
Stimulatory
effect of thyrotropin (TSH) on interleukin 2 (IL2) release from human peripheral
blood lymphocytes and dose-response study in vitro.
Horm.
Metab. Res. 1993, Nov; 25(11); 598-9.
12.
ADRIAANSE R. et al.
Pulsatile
thyrotropin and prolactin secretion in a patient with an mixed thyrotropin and
prolactin secreting pituitary adenoma.
Eur.
J. Endocrinol. 1994, Feb; 130(2); 113-20.
XVI.
SUMMARY OF THE PROTOCOL
|
|
CALIBRATORS
(µl)
|
SAMPLE(S)
CONTROLS
(µl)
|
|
Calibrators
(0-6)
Controls,
Samples
Anti-TSH-HRP
conjugate
|
100
-
50
|
-
100
50
|
|
Incubate
for 1 hour at room temperature with continuous shaking at 700 rpm.
Aspirate
the contents of each well.
Wash
3 times with 400 µl of Wash Solution and aspirate.
|
|
Revelation
Solution
|
200
|
200
|
|
Incubate
for 15 min at room temperature with continuous shaking at 700 rpm.
|
|
Stop
Solution
|
50
|
50
|
|
Read
on a microtiterplate reader and record the absorbance of each well at
450 nm (versus 630 or 650 nm) and 490 nm (versus 630 or 650 nm)
|