|
|
|
T4-RIA-CT
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Reagents | 100 tests kit | colour code | reconstitution |
| Coated tubes with GAM (Goat anti-mouse) | 2x50
|
ready to use
|
|
| Tracer: 125Iodine labelled T4 (HPLC grade) in Phosphate buffer with Bovine casein and azide (<0.1%) | 1vial 21 ml 111kBq | red | ready to use |
| standard zero in human serum with gentamycin and thymol | 1 vial lyoph. | yellow | add 0.5 ml distilled water |
| standard 1 to 5 (see exact values on labels) in human serum with gentamycin and thymol | 5 vials lyoph. | yellow | add 0.5 ml distilled water |
| Ab: anti-T4 (monoclonal) antibodies in phosphate buffer with BSA and thymol | 1 vial lyoph. | blue | add 11ml distilled water |
| wash solution conc. (tris-HCl) | 1 vial 10 ml | brown | dilute 70x with distilled water (use a magnetic stirrer) |
| control 1&2 in human serum with gentamycin and thymol | 2 vials lyoph. | silver | add 0.5 ml distilled water |
5.
SUPPLIES NOT PROVIDED
The
following material is required but not provided in the kit:
1.
Distilled water
2.
Pipettes for delivery of: 20 μl, 100 µl, 200 µl and 500 μl
(the use of accurate pipettes with disposable plastic tips is recommended)
3.
Vortex mixer
4.
Magnetic stirrer
5.
Tube shaker
6.
5 ml automatic syringe (Cornwall type) for washing
7.
Aspiration system (optional)
8. Any gamma counter capable of measuring 125I may be used (minimal yield 70%).
6.
REAGENT PREPARATION
a.
Calibrators:
Reconstitute
the calibrators with 0.5 ml distilled water.
b.
Controls:
Reconstitute the controls with 0.5 ml distilled water.
c.
Anti-T4:
Reconstitute
the anti-T4 with 11 ml distilled water.
d. Working Wash solution: Prepare an adequate volume of Working Wash solution by adding 69 volumes of distilled water to 1 volume of Wash Solution (70x). Use a magnetic stirrer to homogenize. Discard unused Working Wash solution at the end of the day.
7.
STORAGE AND
EXPIRATION DATING
OF REAGENTS
-
Before opening or reconstitution, all kits components are stable until
the expiry date, indicated on the label, if kept at 2 to 8°C.
-
After reconstitution, calibrators and controls are stable for 7 days at
2-8°C.
For longer storage periods, aliquots should be made and kept at –20°C
for maximum 3 months. Avoid
subsequent freeze-thaw cycles.
-
After reconstitution, the anti-T4 antibodies are stable for 6 weeks at
2-8°C. DO
NOT
FREEZE.
-
Freshly prepared Working Wash solution should be used on the same day.
- After its first use, tracer
is stable until expiry date, if kept in the original well-closed vial at 2 to 8°C.
-
Alterations in physical appearance of kit reagents may indicate
instability or deterioration.
8.
SPECIMEN COLLECTION
AND PREPARATION
-
Serum or plasma samples must be kept at 2‑8°C.
-
If the test is not run within 24 hrs, storage at ‑20°C is
recommended.
-
Avoid subsequent freeze-thaw cycles.
-
Serum or Plasma (EDTA or heparin) provides similar results.
9.
PROCEDURE
A.
Handling notes
Do
not use the kit or components beyond expiry date.
Do
not mix materials from different kit lots.
Bring
all the reagents to room temperature prior to use.
Thoroughly
mix all reagents and samples by gentle agitation or swirling.
Use
a clean disposable pipette tip for addition of each different reagent and sample
in order to avoid cross-contamination. High
precision pipettes or automated pipetting equipment will improve the precision.
Respect
the incubation times.
Prepare
a calibration curve for each run, do not use data from previous runs.
B.
Procedure
1.
Label coated tubes in duplicate for each calibrator, control and sample.
For the
determination of total counts, label 2 normal tubes
2.
Briefly vortex calibrators, controls and samples and dispense 20μl
of each into the respective tubes.
3.
Dispense 200 µl of 125Iodine labelled T4 into each tube,
including the uncoated tubes for total counts.
4.
Dispense 100 µl of anti-T4 into each tube, except tubes for total
counts.
5.
Shake the tube rack gently by hand to liberate any trapped air bubbles.
6.
Incubate for 1 hour at room temperature with continous shaking.
7.
Aspirate (or decant) the content of each tube (except total counts).
Be sure that the plastic tip of the aspirator reaches the bottom of the
coated tube in order to remove all the liquid.
8.
Wash tubes with 2 ml Working Wash solution (except total counts) and
aspirate (or decant). Avoid foaming during the addition of the Working Wash
solution.
9.
Let the tubes stand upright for two minutes and aspirate the remaining
drop of liquid.
10.
Count tubes in a gamma counter for 60 seconds.
10.
CALCULATION OF
RESULTS
1.
Calculate the mean of duplicate determinations.
2.
Calculate the bound radioactivity as a percentage of the binding
determined at the zero calibrator point (0) according to the following formula :
3.
Using a 3 cycle semi-logarithmic or logit‑log graph paper, plot the
(B/B0(%)) values for each calibrator point as a function of the T4 concentration
of each calibrator point. Reject
obvious outliers.
4.
Computer assisted methods can also be used to construct the calibration
curve. If automatic result processing is used, a 4-parameter logistic function
curve fitting is recommended.
5.
By interpolation of the sample (B/B0 (%)) values, determine the T4
concentrations of the samples from the calibration curve.
6.
For each assay, the percentage of total tracer bound in the absence of
unlabelled T4 (B0/T) must be checked.
11.
TYPICAL DATA
The
following data are for illustration only and should never be used instead of the
real time calibration curve.
T4 |
cpm |
B/Bo
(%) |
|
Total
count |
32598 |
B/B0 |
|
Calibrator
0
nmol/l 12.8
nmol/l 32
nmol/l 80
nmol/l 200
nmol/l 500
nmol/l |
12661
11434
8460
4554
2020
932
|
100.0
90.3
66.8
36.0
16.0
7.4
|
12.
PERFORMANCE AND
LIMITATIONS
A.
Detection limit
Twenty
zero calibrators were assayed along with a set of other calibrators.
The detection limit, defined as the apparent concentration two standard
deviations below the average counts at zero binding, was < 5 nmol/l.
B.
Specificity
The
percentage of cross-reaction estimated by comparison of the concentration
yielding a 50% inhibition are respectively:
|
Compound |
Cross-Reactivity (%) |
|
L-
thyoxine (L-T4) D-thyroxine
(D-T4) L-3,3’,5
- triiodothyronine (L-T3) L-3,3’,5’
- triiodothyronine (rT3) |
100
48
1.01
7 |
Note:
this table shows the cross-reactivity for the anti T4
C.
Precision
|
INTRA-ASSAY
PRECISION |
INTER-ASSAY
PRECISION |
||||||
|
Serum |
N |
<
>
± SD (nmol/l) |
CV (%) |
Serum |
N |
<
>
± SD (nmol/l) |
CV (%) |
|
A B |
10 10 |
32.4
± 1.8 183.8
± 5.9 |
5.6 3.2
|
A B |
18 20 |
32.7
± 2.1 235.3
± 15.2 |
6.5 6.5
|
SD:
Standard Deviation; CV: Coefficient of variation
D.
Accuracy
DILUTION TEST
|
Sample |
Dilution |
Theoretical
Concent. (nmol/l) |
Measured
Concent. (nmol/l) |
|
A |
1/1 1/2 1/4 1/8 1/16 |
-
212.5
106.3
53.1
26.6 |
425
194.3
89.1
48.8
22.5 |
Samples
were diluted with zero calibrator.
RECOVERY
TEST
|
Sample |
added
T4 (nmol/l) |
Recovered
T4 (nmol/l) |
Recovered (%) |
|
1 |
32.2
64.4
128.7
257.4
386.1
|
27.1
67.8 132.6 290.5 444.8 |
84%
105%
103%
113%
115%
|
To
the best of our knowledge, no international reference material exists for this
parameter.
E.
Time delay between last calibrator and sample dispensing
As
shown hereafter, assay results remain accurate even when a sample is dispensed
60 minutes after the calibrator has been added to coated tubes.
TIME DELAY
|
Serum nmol/l |
0' |
20' |
40' |
60' |
|
C
1 C
2 |
46.5 202.5 |
43.0 212.6 |
38.6 207.3 |
36.4 201.0 |
13.
INTERNAL QUALITY
CONTROL
- If the results
obtained for Control 1 and/or Control 2 are not within the range specified on
the vial label, the results cannot be used unless a satisfactory explanation for
the discrepancy has been given.
-
If desirable, each laboratory can make its own pools of control samples,
which should be kept frozen in aliquots.
-
Acceptance criteria for the difference between the duplicate results of
the samples should rely on Good Laboratory Practises.
14.
REFERENCE INTERVALS
These
values are given only for guidance; each laboratory should establish its own
normal range of values.
T4
concentrations for untreated euthyroid subjects ranged from 64 to 158 nmol/l
(n=80).
15.
PRECAUTIONS AND
WARNINGS
Safety
For
in vitro diagnostic use only.
This
radioactive product can be transferred to and used only by authorized persons;
purchase, storage, use and exchange of radioactive products are subject to the
legislation of the end user's country. In
no case the product must be administered to humans or animals.
All
radioactive handling should be executed in a designated area. away from regular
passage. A logbook for receipt and
storage of radioactive materials must be kept in the lab.
Laboratory equipment and glassware, which could be contaminated with
radioactive substances, should be segregated to prevent cross contamination of
different radioisotopes.
Any
radioactive spills must be cleaned immediately in accordance with the radiation
safety procedures. The radioactive
waste must be disposed of following the local regulations and guidelines of the
authorities holding jurisdiction over the laboratory. Adherence to the basic rules of radiation safety provides
adequate protection.
The
human blood components included in this kit have been tested by European
approved and/or FDA approved methods and found negative for HbsAg, anti-HCV,
anti-HIV-1 and 2. No known method
can offer complete assurance that human blood derivatives will not transmit
hepatitis, AIDS or other infections. Therefore,
handling of reagents, serum or plasma specimens should be in accordance with
local safety procedures.
All
animal products and derivatives have been collected from healthy animals. Bovine
components originate from countries where BSE has not been reported.
Nevertheless, components containing animal substances should be treated as
potentially infectious.
Avoid
any skin contact with reagents (sodium azide as preservative).
Azide in this kit may react with lead and copper in the plumbing and in
this way form highly explosive metal azides.
During the washing step, flush the drain with a large amount of water to
prevent azide build-up.
Do
not smoke, drink, eat or apply cosmetics in the working area.
Do not pipette by mouth. Use
protective clothing and disposable gloves.
16. BIBLIOGRAPHY
1.
Davis J.P. (1991)
Cellullar
actions of thyroid hormones.
The
Thyroid. 190-230. 6th
ed. Braverman L.E., Ultiger R.D.
2.
Stockigt J.R. (1996)
Guidelines
for diagnosis and monitoring of thyroid disease: nonthyroidal illness.
Clin.
Chem. 42 (1); 188-192.
3.
Fisher D.A. (1996)
Physiological
variations in thyroid hormones: physiological and pathophysiological
considerations.
Clin.
Chem. 42 (1); 135-139.
4.
Nelson J.C. and Wilcox R.B. (1996)
Analytical
performance of free and total thyroxine assays.
Clin.
Chem. 42 (1); 146-154.
5.
Robbins, J. (1973)
Radioassay and Thyroid Gland.
Metabolism,
22 (8) : 1021.
|