|
|
|
DHEA-SO4
ELISA
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Calibrator |
Optical Units |
|
Calibrator
0 (0 mg/ml) |
1.92 |
|
Calibrator
1 (0,1 mg/ml) |
1.53 |
|
Calibrator
2 (0,5 mg/ml) |
1.17 |
|
Calibrator
3 (1 mg/ml) |
0.88 |
|
Calibrator
4 (2,5 mg/ml) |
0.59 |
|
Calibrator
5 (5 mg/ml) |
0.35 |
|
Calibrator
6 (10 mg/ml) |
0.17 |
It
is strongly recommended that each laboratory should determine its own normal and
abnormal values.
Premenopausal: 0.8 - 3.9 mg/ml
Term Pregnancy: 0.2 - 1.2 mg/ml
Postmenopausal: 0.1 - 0.6 mg/ml
Normal Men
Newborns (both sexes)
Conversion
factor: 1 mg/ml
= 2.6 mmol/L
The
following materials have been checked for cross reactivity. The percentage
indicate cross reactivity at 50% displacement compared to DHEA-S.
Steroid
% Cross Reaction
DHEA-S
100.0
Androstenedione
20.9
Androsterone
8.5
Androsterone-Sulfate
<
0.1
Progesterone
4.7
Testosterone
0.3
Estrone
0.9
Estriol
< 0.1
Estradiol
< 0.1
Estradiol
Sulfate
< 0.1
Cortisol
<
0.1
7.3 Sensitivity
The
lowest detectable level of DHEA-S -defined as the DHEA-S concentration given by
the mean absorbance of the zero calibrator minus 2 standard deviations- was
assessed to be approx. 0.02 mg/ml.
It
is recommended to use control samples according to state and federal
regulations. The use of control samples is advised to assure the day to day
validity of results. Use controls at both normal and pathological levels.
The
controls and the corresponding results of the QC-Laboratory are stated in the QC
certificate added to the kit. The values stated on the QC sheet always refer to
the current kit lot and should be used for direct comparison of the results.
It
is also recommended to make use of national or international Quality Assessment
programs in order to ensure the accuracy of the results.
Employ
appropriate statistical methods for analysing control values and trends. If the
results of the assay do not fit to the established acceptable ranges of control
materials patient results should be considered invalid.
In
this case, please check the following technical areas: Pipetting and timing
devices; photometer, expiration dates of reagents, storage and incubation
conditions, aspiration and washing methods.
Intra
and Inter Assay Variation
The within assay variability is shown below:
Intraassay
Interassay
|
Serum |
n |
<X>
±
SD µg/ml |
CV % |
n |
<X> ± SD µg/ml |
CV % |
|
1 |
20 |
0.22
±
0.01 |
5.07 |
21 |
0.24
±
0.02 |
8.67 |
|
2 |
21 |
1.31
±
0.03 |
2.13 |
21 |
1.85
±
0.05 |
2.96 |
|
3 |
21 |
3.93
±
0.12 |
3.10 |
21 |
4.26
±
0.16 |
3.66 |
|
Serum |
Endogenous DHEA-S (µg/ml) |
Added DHEA-S (µg/ml) |
Measured conc. (µg/ml) |
Recovery % |
|
1 |
2.12 |
0.5 1.0 2.0 |
2.45 3.11 4.24 |
93.4 99.6 102.9 |
|
2 |
0.38 |
0.5 1.0 4.0 |
0.87 1.35 4.44 |
99.1 97.8 101.5 |
|
Serum |
Dilution factor |
Measured conc. mg/ml |
Recovery % |
|
1 |
Undiluted 1:2 1:4 1:8 1:16 |
2,82 1.28 0.76 0.34 0.16 |
91.0 108.2 95.9 91.9 |
|
2 |
Undiluted 1:2 1:4 1:8 1:16 |
4.71 2.25 1.06 0.59 0.31 |
95.5 90.4 100.4 105.7 |
|
3 |
Undiluted 1:2 1:4 1:8 1:16 |
7.45 3.75 1.86 0.93 0.46 |
100.6 100.0 99.4 98.8 |
Any
improper handling of samples or modification of this test might influence the
results. Interferences caused by improper sample handling are explained in the
chapters ‘Specimen - Collection’.
No
hook effect was observed in this test.
The
test must be performed exactly as per the manufacturer’s instructions for use.
Moreover the user must strictly adhere to the rules of GLP (Good Laboratory
Practice) or other applicable national standards and/or laws. This is especially
relevant for the use of control reagents. It is important to always include,
within the test procedure, a sufficient number of controls for validating the
accuracy and precision of the test.
The test
results are valid only if all controls are within the specified ranges and if
all other test parameters are also within the given assay specifications.
9.2 Therapeutical Consequences
Therapeutical
consequences should never be based on laboratory results alone even if all test
results are in agreement with the items as stated under point 9.1. Any
laboratory result is only a part of the total clinical picture of a patient.
Only
in cases where the laboratory results are in acceptable agreement with the
overall clinical picture of the patient should therapeutical consequences be
derived.
The test result
itself should never be the sole determinant for deriving any therapeutical
consequences.
9.3 Liability
Any
modification of the test kit and/or exchange or mixture of any components of
different lots from one test kit to another could negatively affect the intended
results and validity of the overall test. Such modification and/or exchanges
invalidate any claim for replacement.
Claims
submitted due to customer misinterpretation of laboratory results subject to
point 9.2. are also invalid. Regardless, in the event of any claim, the
manufacturer’s liability is not to exceed the value of the test kit. Any
damage caused to the test kit during transportation is not subject to the
liability of the manufacturer.
(1)
Tietz, N. W., Textbook of Clinical Chemistry, Saunders, 1968
11
DHEA-S Flow chart
|
|
Standards |
Sample(s) |
|
Standards
(0-6) ml Controls/Samples
ml Anti-DHEA-S-HRP
conjugate ml |
25 - 200 |
- 25 200 |
|
Mix
and Incubate 30 min. at RT Wash 3 x 400 ml |
||
|
Substrate
solution |
100 |
100 |
|
Incubate 15 minutes at RT |
||
|
Stop
solution |
50 |
50 |
|
Read:
450 nm |
||
|